Surgery X

Category: Brands: . Woodpecker
  • Description
  • Specifications

Product Description

SURGERY X  “Third Generation Woodpecker Surgery”

Clinical Application :

  • Minimal invasive tooth extraction
  • Intralift
  • Sinus Lift
  • Alveoloplasty
  • Apical cyst extraction
  • Crown extension
  • Crest splitting

MICRON CUTTING

  • The vibration amplitude is between 20-100um. Minimum surgical incision is 3.5mm long, 0.5mm wide, which help with small operating area and fast postoperative recovery.

SELECTIVE CUTTING WITH NO INJURY TO SOFT TISSUE

  • For soft tissues (neural, blood vessels and mucous membranes, etc.) there is no cutting effect, which can protect the vascular nerve of the operation area to a large extent, greatly reducing the risk of injuring soft tissues (neural, vascular, mucosal etc. )

COLD CUTTING 

  • Ultrasonic technology produces very little heat while cutting. Iced water can quickly cool the tip and ensure the temperature below 38 Degree Celsius.
  • At the same time, the water spray flushes the wound during
  • operation, bringing no bleeding and enabling clear vision of operation area which makes it easy to operate.

WORKING MODES:

1. Cutting

According to bone density, adjust power and water volume, power levels are divided into 7 levels:

01 – Level 5, 6 and 7 : Class I bone ( for bone with high bone density / cortical bone )

02 – Level 3 and 4 : Class II (for bone with high bone density / large cortical bone )

03 – Level 2 and 3 : Class II ( for bone w/ uniform bone density / good cortical bone )

04 – Level 1 and 2 : Porotic Bone ( bone with low density / poor cortical bone )

2. Periodontal and Endodontic Mode

Under teeth mode, the power and water volume are adjustable.

3. Cleaning Mode

Step on foot pedal for 3sec to confirm cleaning, then the main unit can automatically complete pipr cleaning for 25sec. After each treatment, please clean the waterway (including pump line and handpiece)

 

 

 

 

TECHINICAL DATA:

  • Software version: BoneSurgical3-V1.0.0
  • Device for intermittent operation: 60s ON, 10s OFF
  • Power-supply voltage: ~100V-240V 50Hz/60Hz 150VA
  • Fuses: 2×1.6AT 250V
  • Working frequency: 24kHz~36kHz
  • Dimension: 276mm x 267mm x 110mm
  • Weight: 2.8kg
  • Water Flow: 25~110ml/m
  • Applied parts: Handpiece and Tips Protection systems and tripping time of the APC:
    No handpiece connected: 10ms
    Cord interrupted: 10ms
  • Tips broken or not correctly tightened: <500ms
  • Protection by discharge to earth: 10ms
  • Alarm: Front display show the e (see point 7.3 and 14)
  • Operation environment:
    a) Environment temperatuer: +5°C~+40°C
    b) Relative humidity: 30%~75%
    c) Atmosphere pressure: 70kPa~106kPa
    d) Temperature in the water inlet of water-cooling equipment is not higher than
    25°C.
  • Delivery and store environment: This equipment should be stored in a
    room where the relative humidity is 10% ~ 93%, atmospheric pressure is 70kPa
    to106kPa, and the temperature is -20℃ ~ +55℃.

CONFIGURATIONS:

  • Handpiece x 1
  • Torque Wrench x 1
  • Tip x 7
  • Tube for Peristatic Pump x 3
  • Foot Pedal